Radiation Dose Calculation and Fetal Risk Assessment During Mammography in Pregnancy Abstract

  Share :          
  79

Abstract Pregnancy-associated breast cancer is relatively rare, but its incidence is rising as the age at childbearing increases. When a pregnant woman presents with a breast mass or suspicious symptoms, mammography raises a legitimate concern about fetal safety. This article reviews the physical basis of fetal dose in mammography, the methods used to calculate and measure it, and how the associated risks are assessed. It concludes that fetal dose from standard mammography is extremely small, far below the thresholds for known harmful effects. 1. Introduction Breast cancer occurs in roughly one in every 3,000 to 10,000 pregnancies. Diagnosis is often delayed because physiological changes in the breast during pregnancy and lactation (increased density and engorgement) mask symptoms and make clinical examination difficult. Diagnostic imaging therefore remains important. Ultrasound is the first-line modality, followed by mammography when needed. The practical question is: how much dose does the fetus actually receive, and is it clinically significant? 2. Physical Basis of Fetal Dose Mammography uses low-energy X-rays (about 25 to 35 kVp) directed at the breast only, with tight collimation. The fetus, located far from the primary beam, receives dose through three pathways: Internal scatter within the patient: Photons scattered in the breast and chest. This falls off rapidly with distance and is attenuated by abdominal and pelvic tissue. Tube housing leakage: Very limited in equipment that meets standards. Scatter from equipment and surroundings: A negligible contribution. Fetal dose rises slightly later in pregnancy because the uterus rises closer to the chest, but it remains very small. 3. Dose Quantities Entrance Surface Air Kerma (ESAK): Measured at the breast surface. Average Glandular Dose (AGD): The reference quantity for breast risk, calculated by the Dance method: AGD = K × g × c × s where K is the incident air kerma at the breast surface, g is a conversion factor depending on compressed breast thickness, c corrects for glandular fraction, and s corrects for the target/filter combination. Fetal absorbed dose: Expressed in gray (Gy) or milligray (mGy), and often in microgray (µGy) because of its small size. 4. Methods for Calculating and Measuring Fetal Dose Direct measurement on phantoms: Thermoluminescent dosimeters (TLD) or optically stimulated luminescence (OSL) dosimeters are placed in the uterine position of an anthropomorphic phantom, and imaging is performed with routine clinical parameters. Dosimeter positions can be varied to represent different stages of pregnancy. Monte Carlo simulation: Codes such as MCNP, GEANT4 and EGSnrc are used with computational pregnant-woman phantoms representing different gestational months, tracking photon histories to calculate dose. This exposes no one to radiation and allows many variables to be studied. Estimation using conversion coefficients: The measured surface dose is multiplied by a published conversion coefficient: Fetal dose ≈ total ESAK × conversion coefficient Simplified illustrative example: If the total surface dose for four views is about 12 mGy and the conversion coefficient to the uterus is on the order of 10⁻⁴ to 10⁻³, the estimated fetal dose ranges from about 1 to 12 µGy (0.001 to 0.012 mGy). These numbers only illustrate the method. Local equipment measurements and published coefficients should be used in practice. Equipment indicators: Modern units display dose values automatically, and these can be recorded in the patient's file for documentation. 5. Published Values Fetal dose from a standard bilateral mammogram (four views) generally falls between roughly 0.001 and 0.01 mGy. Source Approximate dose Bilateral mammography (fetal dose) 0.001 – 0.01 mGy Natural background radiation over a pregnancy (~9 months) ~1 – 2 mGy Threshold for possible deterministic effects ~50 – 100 mGy and above Mammography dose is thus a tiny fraction of the natural background the fetus receives, and thousands of times below the threshold for harm. 6. Fetal Radiation Risk Assessment 6.1 Deterministic effects These effects have a dose threshold and include pregnancy loss, congenital malformations, growth retardation and neurological effects. According to ICRP Publication 84, doses below about 100 mGy are not a reason to terminate a pregnancy. The central nervous system is most sensitive between roughly the 8th and 15th weeks of gestation. Because mammography doses are four to five orders of magnitude below these thresholds, the likelihood of such effects is practically zero. 6.2 Stochastic effects These effects are assumed to have no threshold, and the linear no-threshold (LNT) model is used for radiation protection purposes. The main concern is childhood cancer after in utero exposure. Commonly cited estimates suggest an increase of roughly one case per about 1,700 children for every 10 mGy, with considerable uncertainty. The baseline risk of childhood cancer with no exposure is about 0.2 to 0.3%. Applying this to a dose of 0.005 mGy gives an additional risk on the order of one in several million, which is statistically indistinguishable from baseline risk. 6.3 Maternal breast risk The mother receives a mean glandular dose of about 1 to 2 mGy per view. This risk is small in the context of diagnosing a suspicious finding and is weighed against the benefit of early detection. 7. Professional Recommendations Major professional bodies, including ACOG, ACR and ICRP, agree on the following: Medically necessary diagnostic imaging should not be withheld because of pregnancy alone. Ultrasound is the first-line modality for evaluating a breast mass in a pregnant patient. Mammography is used when there is doubt, or when calcifications need evaluation or disease needs staging. Routine screening mammography is not recommended for asymptomatic pregnant women unless there are special indications. The examination is performed with the lowest reasonably achievable dose (the ALARA principle). The rationale and dose are documented in the medical record. 8. Practical Dose-Reduction Measures Justification: Confirm that the examination will change management and that ultrasound is insufficient. Necessary views only: Limit imaging to the essential craniocaudal (CC) and mediolateral oblique (MLO) views of the affected side, adding views only when needed. Tight collimation: Confine the beam to the breast. Abdominal lead shielding: Some centers use it to reassure the patient. Because the dose is already tiny and the relevant radiation is internally scattered, it is not necessary and must not interfere with image quality or automatic exposure control. Parameter optimization: Use automatic exposure control and appropriate compression to avoid repeat exposures. Documentation and review: Record the dose and review it periodically with the medical physicist. 9. Patient Communication and Informed Consent Clear explanation is a core part of care. The team should explain in plain language that the fetal dose is far lower than everyday natural radiation, that the risk of delayed breast cancer diagnosis outweighs the risk of the examination, and that precautions are in place. Realistic comparisons (for example, equating the dose to days of natural background radiation) are more useful than bare numbers. Informed consent should be obtained and documented. Pregnancy status should also be verified before any X-ray examination of women of childbearing age, according to institutional procedures. 10. Conclusion Physical and experimental data show that fetal dose from standard mammography amounts to thousandths of a milligray. This is far below the thresholds for deterministic effects and carries an additional stochastic risk that cannot practically be distinguished from baseline risk. When there are clear clinical indications, mammography is therefore a relatively safe examination during pregnancy, provided that ALARA is applied and ultrasound is used first. Dose calculation and documentation, carried out jointly by radiology professionals and the medical physicist, remain an essential part of quality assurance and patient reassurance.